The Healer Infrared Laser\n 810\/830nm, 2000mW\nUS FDA Cleared\nSafety and electrical testing completed by Intertek\nIRB approved <\/p>\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\t\t
\n\t\t\t\t\t\t\tThe functionality and product capabilities and coverage of the Healer laser is not\ncomprehensive. Please elaborate\t\t\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\t\t
\n\t\t\t\t
\n\t\t\t\t\t\t\tThe laser works in continuous wave mode. It was a built-in timer that sets treatments in 30\nsecond intervals or you can use it just using the start and stop button. The device automatically\nshuts off when the set time is reached and a manual mode where treatment continues until the\ndevice is manually shut off. The unit is powered down by pressing the power button (battery\nmode). None of these functions are controlled by software. The software controls only the timer\nand time display. Software is embedded into a single chip circuit which contains the CPU and\nmemory. These chips are used in simple circuits. The device can maintain skin temperature of\n40-45 C for at least 10 minutes based on the study conducted on three live human subjects. \t\t\t\t\t\t<\/div>\n\t\t\t\t<\/div>\n\t\t\t\t\t<\/div>\n\t\t<\/div>\n\t\t\t\t\t<\/div>\n\t\t<\/section>\n\t\t\t\t\n\t\t\t\t\t\t